For the units running on Microsoft Windows Server 2008-R2, Windows XP, and Windows 7, the letter instructs that the devices are not to be networked on these systems and that a software patch cannot be made available to these systems since the operating systems are not supported by Microsoft.Ī second cover letter and customer notification were issued dated updating the second safety instruction regarding the units running on Microsoft Windows Server 2008-R2, Windows XP, and Windows 7 saying that these systems must be rebooted before manually exporting tests. For systems that are on Windows 8.1 and above, the issue will be corrected through a software patch. The recalling firm issued letters dated beginning. Cardiosoft Software pdf#If a certain sequence of events occur, the pdf test report that belongs to one patient will appear in a different patient's record when viewed in an Electronic Medical Record (EMR) or a similar system. Class 2 Device Recall Case Cardiac Assessment System and CardioSoft Cardiac Testing SystemĬomputer, diagnostic, programmable - Product Code DQKĬase Cardiac Assessment System for Exercise Testing and CardioSoft Diagnostic System Exercise Stress Testing ECG applicationĬASE and CardioSoft - v6.6x, v6.7x and v7.0